Ateq D670
  • Leak & Flow
  • Leak & Flow
  • Leak & Flow

Ateq D670

After 40 years of experience and 100 000 leak detectors and flow testers installed worldwide, ATEQ presents its revolutionary flow tester operating on the principle of direct or indirect flow measurement.

The ATEQ D670 is equipped with a large 3U 19-inch rack case, which allows installation of additional elements inside the instrument, including valves for measurement channels switching or external pressure exhaust for contaminated parts.

ATEQ D670 is a new Series 6 generation of flow testers with:

  • new design differential pressure transducers
  • new electronic and pneumatic modules
  • new user-friendly interface
  • new accessories and functions

The innovative technology of the Series 6 instruments was created to provide customers with industrial devices that meet the demand for quick and reliable flow testing as well as user-friendly operation.

  • Graphic user-friendly interface with high-contrast color screen
  • Direct or indirect leak and flow measurement
  • Various flow measurement ranges: from 1 l/h to 10.000 l/h
  • Flow measurement in:
    • SI standard units
    • US units
    • custom calibrated units
  • The entire measuring system integrated in one module
  • Possibility of programming 128 test programs
  • Communication:
    • RS232: printer, barcode reader, PC computer
    • USB «master»: transfer to external USB memory (parameters, results, statistics, …)
    • USB «slave»: supervision
    • Ethernet/IP or USB «master» on the front panel or SD reader for saving results.
    • Available communication protocols: ModBus, ProfiBus, DeviceNet, ProfiNet, EtherNet/IP, EtherCAT.
  • 9 inputs / 7 outputs board – enables selection of 128 programs, test cycle START/STOP, OK/NOK result signals, etc.
  • Date and time indication, language and unit selection, ability to adapt tests to your requirements, automatic measurement cycle, …
  • Electronic regulator for automatic test pressure adjustment
  • 2 pneumatic outputs for automation control
  • Connections to the test item
  • Programmable 6 x 24 V outputs for external devices
  • 2 analogue outputs 0-10V or 4-20mA
  • Automatic calibration check mode
  • Tight shut-off and switching valves for construction of complex test circuits
  • Pressure and flow calibrator

Flow testing:
gas cookers, stoves, valves, gas taps, heaters, hydrophobic water-repellent materials and membranes, catheters, dialysis devices, …

Leak detection:
valves, gearboxes, complete engines and components, …

Functional testing:
aerosol sprays, injection systems, …

FLOW MEASUREMENT
RANGE ACCURACY
1 l/h ± ( 2.5% F + 0.001 l/h )
5 l/h ± ( 2.5% F + 0.005 l/h )
30 l/h ± ( 2.5% F + 0.03 l/h )
80 l/h ± ( 2.5% F + 0.08 l/h )
150 l/h ± ( 2.5% F + 0.15 l/h )
500 l/h ± ( 2.5% F + 0.5 l/h )
1 500 l/h ± ( 2.5% F + 1.5 l/h )
4 000 l/h ± ( 2.5% F + 4 l/h )
10 000 l/h ± ( 2.5% F + 10 l/h )
*F – measured flow value
PRESSURE MEASUREMENT
RANGE ACCURACY
Pmax = 75 mbar ± ( 1.5% P + 0.2 mbar )
P < 0,3 bar ± ( 1.5% P + 1 mbar )
0,3 < P < 1 bar ± ( 1.5% P + 3 mbar )
1 < P < 5 bar ± ( 1.5% P + 7.5 mbar )
5 < P < 10 bar ± ( 1.5% P + 15 mbar )
*P – measured pressure value

MECHANICAL PRESSURE REGULATOR
from -800 to -20 mbar • from 8 to 60 mbar • from 5 to 140 mbar • from 0,1 to 2 bar •
• from 0,4 to 4 bar • from 1 to 10 bar

ELECTRONIC PRESSURE REGULATOR
from 10 to 200 mbar • from 20 to 500 mbar • from 0.2 to 2 bar •
• from 0.5 to 5 bar • from 1 to 10 bar