Ateq ERD620
  • Leak & Flow
  • Leak & Flow
  • Leak & Flow

Ateq ERD620

After 40 years of experience and 100 000 leak detectors and flow testers installed worldwide, ATEQ presents its revolutionary compact flow and pressure event tester.

ATEQ ERD620 is a new Series 6 generation of pneumatic testers with:

  • new design differential pressure transducers
  • new electronic and pneumatic modules
  • new user-friendly interface
  • new accessories and functions

ATEQ ERD620 is used to test components that are expected to react to a specific pressure. The instrument generates an increasing or decreasing pressure ramp until the expected event occurs, such as:

  • relay contacts opening or closing
  • safety valve opening or closing
  • nominal flow achieved at certain pressure

The tester checks whether the expected event occurred within the specified pressure range, while measuring the air flow.

 

  • Graphic user-friendly interface with high-contrast color screen
  • Automatic test pressure adjustment
  • Programmable pressure ramp increase and decrease
  • Direct or indirect flow measurement
  • Various flow measurement ranges: from 1 l/h to 10.000 l/h
  • Flow measurement in:
    • SI standard units
    • US units
    • custom calibrated units
  • The entire measuring system integrated in one module
  • Possibility of programming 128 test programs
  • Communication:
    • RS232: printer, barcode reader, PC computer
    • USB «master»: transfer to external USB memory (parameters, results, statistics, …)
    • USB «slave»: supervision
    • Ethernet/IP or USB «master» on the front panel or SD reader for saving results.
    • Available communication protocols: ModBus, ProfiBus, DeviceNet, ProfiNet, EtherNet/IP, EtherCAT.
  • 9 inputs / 7 outputs board – enables selection of 128 programs, test cycle START/STOP, OK/NOK result signals, etc.
  • Date and time indication, language and unit selection, ability to adapt tests to your requirements, automatic measurement cycle, …
  • 2 pneumatic outputs for automation control
  • Programmable 6 x 24 V outputs for external devices
  • 2 analogue outputs 0-10V or 4-20mA
  • Connections to the test item
  • Automatic calibration check mode
  • Tight shut-off and switching valves for construction of complex test circuits

Flow and pressure event testing of safety valves and pressure safety systems, non-return valves, sensors, pressure transducers and switches, …

FLOW MEASUREMENT
RANGE ACCURACY
1 l/h ± ( 2.5% F + 0.001 l/h )
5 l/h ± ( 2.5% F + 0.005 l/h )
30 l/h ± ( 2.5% F + 0.03 l/h )
80 l/h ± ( 2.5% F + 0.08 l/h )
150 l/h ± ( 2.5% F + 0.15 l/h )
500 l/h ± ( 2.5% F + 0.5 l/h )
1 500 l/h ± ( 2.5% F + 1.5 l/h )
4 000 l/h ± ( 2.5% F + 4 l/h )
10 000 l/h ± ( 2.5% F + 10 l/h )
*F – measured flow value
PRESSURE MEASUREMENT
RANGE ACCURACY
Pmax = 75 mbar ± ( 1.5% P + 0.2 mbar )
P < 0,3 bar ± ( 1.5% P + 1 mbar )
0,3 < P < 1 bar ± ( 1.5% P + 3 mbar )
1 < P < 5 bar ± ( 1.5% P + 7.5 mbar )
5 < P < 10 bar ± ( 1.5% P + 15 mbar )
*P – measured pressure value

ELECTRONIC PRESSURE REGULATOR
from 10 to 200 mbar • from 20 to 500 mbar • from 0.2 to 2 bar •
• from 0.5 to 5 bar • from 1 to 10 bar