Ateq Master Jet

ATEQ MASTER JET – flow master standards (jet, orifice) are used for calibration checking of all types of measuring equipment in accordance with ISO17025 requirements, such as flow testers, flow transducers or rotameters.

The master jets are delivered with a certificate of calibration in a protective box, mounted on a tube fitting.

 

PRESSURE Type of master jet / rate of air flow in dm3/h at 20°C and 1013 mbar
10
(100 µm)
20
(200 µm)
35
(350 µm)
59
(590 µm)
80
(800 µm)
112
(1120 µm)
160
(1600 µm)
20 mbar 0,98 5 17 51 93 180 405
100 mbar 2,64 13 40 117 221 387 922
250 mbar 4,36 21 64 185 354 607 1473
500 mbar 6,54 29 90 262 482 890 1947
1 bar 9,6 43 129 372 676 1262
1,5 bar 12,4 54 162 468 850 1599
2 bar 15 65 194 545 1010 1909
2,5 bar 17,7 76 227 638 1185
3 bar 20,4 88 260 729 1357
3,5 bar 23,4 100 293 819 1524
4 bar 25,2 110 325 910 1694

Note: the values indicated above are given for information only and can vary within ±20%. The true flow value is precisely measured before delivery with an accuracy of ±1.5% of the flow. Special jets can be manufactured within ±5% of its requested value up to 9000 dm3/h.

 

  • Must be used with dry and clean air
  • Must not be used under water and must be put in their box after use
  • Must be periodically checked by ATEQ’s metrology department
  • Checking the zero of the instrument must be done by blanking the instrument test output with a cap
  • An extra jet can be provided as option in the same box to allow double calibration checks or a spare master to be used when one is sent to our laboratory for recertification