Ateq Master Leak

ATEQ MASTER LEAK – leak master standards (orifice) are used for calibration checking of all types of measuring equipment in accordance with ISO17025 requirements, such as leak detectors and pneumatic testers.

The master leaks are delivered with a certificate of calibration in a protective box, mounted on a quick connector.

PRESSURE Type of master leak / rate of air flow
A B 5 C D 50 E F G 1000 5000
20 mbar 8,2 16 34,7 1,3 2,7 5,8 52
50 mbar 8,7 21 41 1,5 3,3 6,8 15 130
150 mbar 6,0 13 27 64 2,2 4,7 10 21 47 383
200 mbar 8,1 17 37 88 3,0 6,3 14 29 64 505
300 mbar 5,8 13 26 57 2,3 4,6 9,8 21 45 99 740
500 mbar 10 22 46 1,7 4,1 8,2 17 38 81 176 1198
750 mbar 17 36 75 2,8 6,7 13 28 62 132 281 1823
850 mbar 19 42 88 3,3 7,8 15 32 73 155 325 2049
1 bar 23 51 1,8 4,0 9,5 19 40 90 190 394 2404
2 bar 58 2,2 4,6 10 24 46 95 234 486 904 4709
4 bar 2,7 6,1 13 27 61 118 233 665 1305 2086 9357
10 bar 11 24 51 99 213 394 729 2810 4620 6143
20 bar 31 65 134 247 512 916 1661
– 0,1 bar 7,8 17 39 78 2,8 6,1 13 28 237
– 0,2 bar 7,1 15 31 75 149 5,3 12 24 53 427
– 0,5 bar 7,4 15 32 69 163 324 11,3 25 52 112 813
– 0,75 bar 9,8 20 42 90 212 418 14,5 29 66 142 991
– 0,85 bar 10,6 22 45 97 225 443 15,4 35 70 149 1032
– 0,95 bar 11,3 23 47 101 235 462 15,9 36 73 153
Rate of air flow at 20°C and 1013 mbar
in (cm3/ h) in (cm3/ min)

Note: the values indicated above are given for information only and can vary within ±20%. The true leak value is precisely measured before delivery with an accuracy of ±1.5% of the flow. Special master leaks can be manufactured within ±5% of its requested value up to 9000 dm3/h.

 

  • Must be used with dry and clean air
  • The O-ring located inside the leak must be checked
  • Must not be used under water and must be put in their box after use
  • The leak can be visually checked by connecting a tube to its outlet and placing the tube’s other end under water to generate bubbles
  • Must be periodically checked by ATEQ’s metrology department
  • Checking the zero of the instrument must be done by blanking the instrument test output with a cap
  • An extra master leak can be provided as option in the same box to allow double calibration checks or a spare master to be used when one is sent to our laboratory for recertification